Monitors — From SpO2 to the Swan-Ganz
Read this lesson without an account. To use narrated cards and save your progress in the app, create a free account or sign in. No payment is required for the free modules. Knowledge practice does not replace approved training or any skills evaluation your certifying body requires.
Monitors — From SpO2 to the Swan-Ganz
Check this point against current official guidance before relying on it.
- Every monitor answers one specific question — learn which.
- Pulse oximetry, capnography, the invasive lines, the nerve stimulator, warming and adjuncts.
- Confusing two monitors is a classic exam trap; we separate them cleanly.
What does pulse oximetry measure, by what physical principle (which two wavelengths), and what does it NOT tell you?
Check this point against current official guidance before relying on it.
- PULSE OXIMETRY (SpO2) measures arterial oxygen saturation non-invasively by shining two wavelengths of light (about 660 nm red and 940 nm infrared) through tissue and comparing pulsatile absorption. It reads saturation, NOT ventilation, and lags behind an acute airway problem. SpO2 = oxygenation via two-wavelength light absorption. It tells you saturation, not whether the patient is ventilating (that is capnography's job).
- Remember: Pulse ox shines two lights — 660 nm red and 940 nm infrared — and reads SATURATION off the pulse. It tells you oxygen on board, never whether the patient is moving air; it lags a blocked airway.
What does waveform capnography show, and why is a transient waveform insufficient to establish tracheal tube placement?
Check this point against current official guidance before relying on it.
- Waveform capnography displays exhaled carbon dioxide over time. Sustained, clinically appropriate exhaled CO2 supports tracheal tube placement; a transient waveform alone is insufficient. Sudden loss of the waveform requires immediate assessment of the airway, breathing circuit, circulation and monitor. Pulse oximetry assesses oxygen saturation and cannot replace this assessment.
- Remember: Look for sustained, appropriate exhaled carbon dioxide. A transient trace is not enough, and a lost trace requires immediate assessment.
Compare a central venous pressure (CVP) line and a pulmonary artery (Swan-Ganz) catheter: what does each measure?
Check this point against current official guidance before relying on it.
- Invasive hemodynamic monitors differ by what they measure: a CENTRAL VENOUS PRESSURE (CVP) line sits in the great veins/right atrium and reflects right-heart filling, while a PULMONARY ARTERY (Swan-Ganz) catheter advances through the right heart to give pulmonary artery pressure, wedge (occlusion) pressure, cardiac output, and mixed venous oxygen (SvO2). CVP = right-sided filling only. PA catheter = deeper data (PA pressure, wedge as a left-filling surrogate, cardiac output, SvO2). Don't confuse the two.
- Remember: A CVP line reads right-heart filling and nothing more; the Swan-Ganz threads on THROUGH the right heart for PA pressure, wedge, cardiac output, and SvO2. One number versus the whole downstream story.
What does an arterial line provide, and why must its transducer be zeroed and leveled at the phlebostatic axis?
Check this point against current official guidance before relying on it.
- An ARTERIAL LINE gives continuous beat-to-beat blood pressure and a port for arterial blood-gas sampling. The transducer must be zeroed to atmosphere and LEVELED at the phlebostatic axis (right-atrial level); a transducer set too low reads falsely high and too high reads falsely low. Art line = continuous BP + ABG access. Leveling matters: below heart level over-reads, above heart level under-reads. Zero to atmosphere first.
- Remember: Zero the art-line to air, then LEVEL it at the phlebostatic axis. Drop the transducer below the heart and it over-reads; lift it above and it under-reads — the column of fluid is doing the lying.
What does train-of-four stimulation assess, and why do four apparently equal twitches not prove recovery?
Check this point against current official guidance before relying on it.
- Train-of-four stimulation helps assess neuromuscular block. Fewer responses generally indicate deeper block, but four twitches that look equal do not rule out residual block. Use quantitative monitoring at the adductor pollicis and confirm a train-of-four ratio of at least 0.9 before extubation. The anesthesia clinician determines reversal; a visual twitch count alone is not clearance.
- Remember: Count describes depth; a measured ratio confirms recovery. Four visible twitches alone are not clearance.
Name the warming devices that fight perioperative hypothermia and several adjunct monitoring/support devices the tech readies.
Check this point against current official guidance before relying on it.
- Temperature and adjunct devices the tech prepares include internal/external temperature probes, fluid/blood warmers and forced-air warming (to prevent perioperative hypothermia), rapid-infusion devices, transesophageal echocardiography (TEE), the intra-aortic balloon pump (IABP), ultrasound, thromboelastography (TEG), and point-of-care testing (glucometer, ABG, hemoglobin/hematocrit). Warming (forced-air, fluid warmers) fights hypothermia; TEE/IABP/ultrasound/TEG/point-of-care extend monitoring and support. Know each device's purpose for setup.
- Remember: The warming shelf fights hypothermia — forced-air blanket, fluid and blood warmers. The rest of the adjunct cart (TEE, IABP, ultrasound, TEG, point-of-care) each answers one question the monitors cannot.
Pulse oximetry estimates arterial oxygen saturation. Capnography displays exhaled CO2; sustained, clinically appropriate exhaled CO2 supports tracheal tube placement.
Check this point against current official guidance before relying on it.
- CVP and pulmonary artery catheters provide different hemodynamic information. Zero and level an arterial-line transducer. Use quantitative train-of-four monitoring to assess recovery.
Which monitor displays exhaled carbon dioxide, and why must the waveform be sustained and clinically appropriate?
- How does positioning an arterial-line transducer below heart level affect the displayed pressure?