The anesthesia machine has a HIGH-pressure system (gas cylinders — a full E-cylinder of oxygen reads about 1900-2200 psig), an INTERMEDIATE-pressure system (hospital pipeline supply, about 50-55 psig), and a LOW-pressure system (flowmeters and vaporizers, just above atmospheric). The tech traces a gas-supply problem by knowing which pressure section it lives in. Three pressure zones: cylinder (high), pipeline (intermediate ~50 psig), flowmeter/vaporizer (low). A cylinder empties as pressure falls; a pipeline problem shows at ~50 psig.
Remember: Trace any gas problem to its pressure zone: the cylinder gauge falling toward empty is HIGH-pressure (a full O2 E-cyl reads about 1900-2200 psig); the wall pipeline sits steady near 50 psig (intermediate); past the flowmeters it is barely above the room. Name the zone, find the fault.